GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 160 Section 10.0 - PACS / Image Display Test Image
a.) Configure the PACS to reject the DX-modality
b.) Send the PACS Test Image 9
c.) Open the PACS Test Image 9
d.) Open the DICOM header for Image#9
• Verify that the VOI-LUT Sequence tag is not present, and
• Verify that the WC=8192/8192/8192 and WW=16383/12288/24575
Record results in Table 6-2, line number 15
e.) The image should appear visually identical to image 6 if CR-fallback is supported.