Page 220 Section 5.0 - Image Quality Issues
Solutions:
1.) Perform Manual bad pixel calibration. See Section 4.0.
2.) Take exposure again to confirm the bad lines disappear.
3.) Perform QAP again.
5.2 Equi-spaced Zebra White Bad Lines Issue
Detector calibration passed but QAP failed, after taking an exposure, Equi-spaced zebra white bad
lines appeared in whole image.