GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 8 - - Troubleshooting Page 221
Root Cause:
Press exposure switch too quickly after exposure count down clock is displayed in screen
during Detector calibration, so detector scan data is lost.
Solutions:
1.) Do not press exposure switch immediately after exposure count down clock is displayed in
screen. You can delay a little seconds to take exposure, only if you can finish the exposure
within the time.
2.) Repeat Detector calibration
3.) Repeat QAP.