GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 64 Section 8.0 - Software Installation
8.2.3 Backup Protocol Data Base
Note: From SUIF/Configuration/General/Application Language, record the language config. This is
because the database only can be restored at the system configured with the same language
environment.
1.) From the Worklist screen, click Utilities icon. The Service Tools screen appears.
2.) Select [Preferences] first and then [Protocols], from the menu.
3.) Insert a blank CD/DVD into the CD tray and close.
4.) Click [Back up].
5.) A message will appear: Press OK to continue with Protocol Database back up.
6.) Press [OK] after you see the message. The protocol database is saved to the CD/DVD.
7.) Remove the CD/DVD from tray and label it as “Protocol Database” (include the current date
and time).
8.) Place the CD/DVD in a safe location for future use.

Table of Contents