continued
HEAT OUTPUT
Stand by
Moderate Use (8
cases / a 10 hour
day)
Typical Use (13
cases / a 10 hour
day)
Maximum Use
(maximum peak
power during ex-
am)
Room Core System kW BTU/hr kW BTU/hr kW BTU/hr kW BTU/hr
Main disconnect
panel PDB
0.60 2216 0.65 2216 0.65 2216 0.65 2216
1kVA Cabinet UPS -
model 9130
0.15 500 0.37 1257 0.37 1257 0.37 1257
Tech Room Total 7.24 24842 12.29 41888 15 51153 19.65 67003
Total for core system 8.26 28323 13.44 45832 16.63 56685 21.68 73925
Room Options (3) Stand by Moderate Use Typical Use Maximum Use
Exam
Room
7 in room B&W TFT
monitors
0.59 2005
Same values as
Stand by
Same values as
Stand by
Same values as
Stand by
In room AW TFT
monitor
0.12 409
Typical injector 0.09 320
LD Monitor 0.5 1706
Exam room total 1.3 4440
Ctrl
Room
AW work station 0.35 1201
2 AW TFT monitors 0.24 818
Printer 0.31 1054
LD Monitor (4) 0.5 1706
Ctrl room total 1.4 4779
Tech
Room
LD Cabinet 1.0 3412
3kVA LDM UPS -
model 9130
0.37 1257
Tech Room Total 1.37 4669
Typical configuration without
fluoro UPS
10.27 35173 15.15 51640 18.33 62493 23.39 79732
Note (1): Air flow requirements 1200 m
3
/h (706 CFM)
Note (2): For more details, consult appropriate pre–installation manual
Environmental Requirements
132
Innova
TM
IGS 6
5750182-1EN 3