The fitter shall install an X-Ray On lamp in the Exam room, and in the control room if necessary, outside
the patient area, so that they are connected to the equipment through connection terminals.
NOTE
Innova systems with 21 & 31 cm detector provides 24 VAC for driving X-Ray ON line.
Hospital or contractor to provide 1.5 mm
2
(AWG 14) wires to connect to the system.
X-Ray ON lamp is active during X-Ray exposures.
WARNING
THE X-RAY ON LAMP MUST BE INSTALLED IN THE EXAM ROOM TO CONFORM TO
INTERVENTIONAL STANDARD IEC/EN 60601-2-43.
SIGNAL INDICATING THE X-RAY ON SHALL BE PERCEPTIBLE BY THE OPERATOR IN
ALL THE LOCATIONS DEFINED FOR THE PERSONNEL WHO MAY RECEIVE
SCATTERED RADIATION.
In the control room, an additional X-Ray on light must be installed if the console (VCIM) indicator cannot
be perceived by all the persons in the control room.