Chapter 2. Equipment Requirements
2.1 System Components
2.1.1 Presentation of the 3 Rooms
The components shall be installed in three different rooms with different constraints: the Exam Room, the
Control Room and the Technical Room.
Exam Room
This is where the patient is situated. It contains the table on which the patient is lying, the table side user
interfaces (TSUI), the frontal and lateral gantries, the exam monitors, and accessories.
Control Room
This room contains user interface and control monitors. No intentional or unintentional contact with the
patient shall occur with the patient in this area.
Technical Room
This room contains electronic cabinets. No intentional or unintentional contact with the patient shall
occur with the patient in this area. It is strongly recommended that this room is separated from the
Control Room, in order to minimize risks of transmission of airborne pathogens. Its construction should
be adapted to minimize ambient noise level; for example the use of glass doors instead of louvered hung
doors.
Equipment Requirements
5750182-1EN 3
Innova
TM
IGS 6
29