Fluoro, 3 kVA & 1 kVA UPS Shipment Duration
NOTE
In the case the material shipped will be subject to different temperature values during
shipment, the maximum shipment duration can be calculated, depending on the duration
of different temperature values application, and using values given in the tables above
Fluoro UPS (CE and UL)
The maximum shipment duration is given in the table for both Fluoro UPS CE and UL. It depends on the
temperature to which the material will be subject during shipment. This limitation is due to the capability
of batteries included in UPS.
Table 1-3 Maximum shipment duration for Fluoro UPS
Temperature Shipment max duration (Weeks)
55°C (131°F) 2
50°C (122F°) 3
40°C (104°F) 6
30°C (86°F) 12
3 kVA UPS (220 V and 110 V) & 1 kVA UPS
The maximum shipment duration is given in the table for both 3 kVA UPS 220 V and 110 V, and 1 kVA UPS.
It depends on the temperature to which the material will be subject during shipment. This limitation is
due to the capability of batteries included in UPS.
Table 1-4 Maximum shipment duration for 3 kVA UPS & 1 kVA UPS
Temperature Shipment max duration (Weeks)
55°C (131°F) 4
50°C (122F°) 7
40°C (104°F) 14
30°C (86°F) 25
General Requirements
28
Innova
TM
IGS 6
5750182-1EN 3