continued
Route Survey
Ref. no. Area: loading dock, doorway, hallway, turn,
elevator, obstruction, etc.
Limitations: loading capacity, height, width,
depth flooring material, radius, etc.
1.4 Product Storage and Handling Requirements
Product Storage and Handling limits
Relative Humidity and Temperature: Refer to Table 1-1 Storage and transport – Relative Humidity and
Temperature on page 25. To obtain relative humidity and temperature requirements for components
not specified in Table 1-1 Storage and transport – Relative Humidity and Temperature on page 25,
refer to the appropriate component Pre–Installation Manual listed in 2.1.5 System Compatibility on page
71.
Atmospheric Pressure: Refer to Table 1-2 Atmospheric Pressure - Storage and transport on page 27.
To obtain altitude and atmospheric pressure requirements for components not specified in Table 1-2
Atmospheric Pressure - Storage and transport on page 27 refer to the appropriate component Pre–
Installation Manual listed in 2.1.5 System Compatibility on page 71.
NOTICE
Avoid extremes in temperatures
Table 1-1 Storage and transport – Relative Humidity and Temperature
SUB-SYSTEM COMPONENTS RELATIVE HUMIDITY (NON-
CONDENSING) (see NOTE (2))
TEMPERATURE
MIN MAX MIN MAX
Frontal Positioner 5% 95% -40°C -40°F +70°C +158°F
Lateral Positioner 5% 95% -40°C -40°F +70°C +158°F
Omega Table 5% 95% -40°C -40°F +70°C +158°F
TSUI (TSSC, Smart box, Smart handle, TPD) 5% 95% -40°C -40°F +70°C +158°F
General Requirements
5750182-1EN 3
Innova
TM
IGS 6
25