Storage and transport – Relative Humidity and Temperature continued
SUB-SYSTEM COMPONENTS RELATIVE HUMIDITY (NON-
CONDENSING) (see NOTE (2))
TEMPERATURE
MIN MAX MIN MAX
Footswitch 10% 100% -20°C -4°F +60°C +140°F
Monitor LCD Eizo 5% 95% -20°C -4°F +60°C +140°F
Monitor LDM 10% 90% -20°C -4°F +55°C +131°F
Cabinets C1 Frt, C1 Lat, C2 10% 100% -40°C -40°F +70°C +158°F
UPS 3 kVA, 1 kVA 5% 90% -20°C -4°F +40°C +104°F
Power Distribution Box (PDB) CE 5% 95% -30°C -22°F +70°C +158°F
Power Distribution Box (PDB) UL 10% 95% -30°C -22°F +70°C +158°F
UL Fluoro UPS cabinet (*) 0% 95% -20°C -4°F +50°C +122°F
CE Fluoro UPS cabinet (*) 0% 95% -20°C -4°F +50°C +122°F
LDM cabinet LD core 5% 95% -40°C -40°F +70°C +158°F
Detector conditioner 5% 95% -40°C -40°F +70°C +158°F
Tube chiller Coolix 4100 10% 100% -40°C -40°F +70°C +158°F
VCIM 5% 95% -40°C -40°F +70°C +158°F
Digital detector Refer to NOTE (1) below
NOTE
(1): The detector should be stored at 10 to 40 °C (50 to 104 °F) and less than or equal to
90% RH in the plastic wrapped shipping box. (This should include two bags of desiccant as
well). The lowest temperature (e.g. 10 ºC (50 °F)) and humidity is preferable. If they are to be
stored outside of their shipping box or in the inner shipping box without plastic wrapping
they should be stored at 20 °C (68 °F) or less and 30% RH or less. In terms of
transportation, do not expose to temperatures below –20 ºC (–4 °F) in its shipping box for
more than 15 hours. The detector will reach the ambient temperature after 20 to 25 hours.
The detector should not be allowed to reach temperatures less than –10 ºC (14 °F) or
irreparable damage to the detectors scintillator will occur. Care must be taken when
removing a detector from a shipping box. If the detector has been subject to cold
temperatures for an extended period the detector in the box should be allowed to sit in
the plastic wrapped box to reach room temperature. This will prevent condensation from
occurring. Condensation on the detector can cause irreparable damage to the electronics.
Storage 10 to 40 ºC (50 to 104 °F); 10 to 90 % RH, 250 day storage transportation –20 to
+60 ºC (–4 to 140 °F) and 10 to 80% RH. The Detector Conditioner is shipped within GE
Healthcare packaging.
General Requirements
26
Innova
TM
IGS 6
5750182-1EN 3