Gantry Noise measurements continued
Measurements Gantry (FRT, LAT, both) Gantry posi-
tions
Peak (db) Mean (db)
First scopie (anode accel-
eration)
FRT 1 80.2 60.3
LAT 2 80.3 59.6
Fluoro
FRT 1 79.7 59.2
LAT 2 79.5 59.1
FRT and LAT 2 79.8 60.2
NOTE
Noise generated by system cabinets all together is 70dB. If cabinets are installed too close
to the exam room (e.g. behind sliding doors) appropriate measures have to be
implemented during the pre-installation to effectively reduce noise in the exam room.
Environmental Requirements
5750182-1EN 3
Innova