The minimum required average luminance Em shall be of 500Lx and minimum color rendering factor Ra
of 80 as per IEC/EN 12464-1 (Light and lighting. Lighting of work places. Indoor work places: Illumination
requirements for indoor workplaces corresponding to assembly of medium size electrical components,
e.g. control panel) for the electrical industry).
Lighting Relay
Innova systems with 21 & 31 cm detector has the ability to control an external relay that applies power
to the room light (dry contacts).
The relay is to be provided by the hospital or contractor.
The wire size to connect to the C2 cabinet is 1.5 mm
2
(AWG 14).
Relay rating is max 600 VAC
Figure 5-1 ROOM LIGHT cabling between C2 cabinet and PDB