EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ S8 R3

GE LOGIQ S8 R3 User Manual

GE LOGIQ S8 R3
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #281 background imageLoading...
Page #281 background image
GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-3 - Gathering Troubleshooting Data (R1 to R4) 7 - 11
7-3-4 Collecting a Screen Capture with Logs (cont’d)
R3.x.x or later: Zip file configuration
7-3-4-1 Marking Log files
If a customer is experiencing issues during operations, the event can be marked and logged by pressing
Alt+1 or Alt+2 when they occur. When Alt+1 or Alt+2 are pressed, the log will register time and help log
analysis when troubleshooting intermittent issues.
Zipped_log_without_protected_info. Zipped_log_including_protected_info