IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3 - 30 Section 3-7 - Power On/Off
12.) If you have not already written the key down you can do so now before moving on.
NOTE: It is important to write down or take a print of the password and recovery key. Store these and
the USB key in a secure location.
The displayed recovery key string shown is an example and not a real recovery key. The
generated key will always be different.
13.)Select Ok to continue. Encryption of data begins. This may take up to 90 minutes the first time. After
the initial encryption there is no long process for daily use.
Figure 3-20 Disk Encryption - Encrypting Window
14.) If the recovery key needs to be changed, perform disk encryption again. Insert the USB flash drive
and select Change key.
Figure 3-21 Disk Encryption - Change Key
After the disk is encrypted, you will be required to unlock the disk at each startup of the system.
The unlock dialog box will appear when the new patient or print button is pushed.

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals