DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips 7 - 83
7-7-20 When the touch panel does not response
From R4 system, the touch panel changes to electrostatic type. If the keys does not respond, clean the
touch panel first. If the gel or somehting remains on the touch panel, it does not reposnd.
7-7-21 Limitation message during AFI Analysis
The limitaion message “Over Depth too high for analysis xx.xx cm, EXIT" displays when you measure
AFI deeper than 20 cm. Measure whithin 20cm.