DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
9 - 6 Section 9-5 - Tools required
Section 9-5
Tools required
9-5-1 Special tools, supplies and equipment
Also refer to 8-2-6 "Tools needed for servicing LOGIQ S8/LOGIQ E8" on page 8-5.
Table 9-2 Overview of requirements for periodic maintenance
Tool Comments
Digital Volt Meter (DVM)
Anti Static Kit Typically includes antistatic-mat, wrist strap and ground cable.
Anti Static Vacuum Cleaner If available on site.
Safety Analyzer
Any calibrated Electrical Safety Analyzer compliant with AAMI/ESI 1993 or
IEC 60601 or AS/NZS 3551
Phantom If available on site.
CD-R/DVD-R Media
DVD+RW Disc Media blank
B/W Printer Cleaning Sheet see printer user manual for requirements
Color Printer Cleaning Sheet see printer user manual for requirements
Disposable Gloves

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4