Section 4-4 - Functional Checks 4 - 33
4-4-13 Basic Measurements
NOTE: The following instructions assume that you first scan the patient and then press Freeze.
4-4-13-1 Check Distance and Tissue Depth Measurement
1.) Press Measure once to display an active caliper.
2.) Move the Trackball to position the active caliper at the start point (distance) or the most anterior
point (tissue depth).
3.) Press Set to fix the start point.
4.) The system fixes the first caliper and displays a second active caliper.
5.) Move the Trackball to position the second active caliper at the end point (distance) or the most
posterior point (tissue depth).
6.) Press Set to complete the measurement. The system displays the distance or tissue depth value in
the measurement results window.
NOTE: Before you complete a measurement:
To toggle between active calipers, press Measure.
To erase the second caliper and the current data measured and start the measurement again, press
Clear once.
NOTE: To rotate through and activate previously fixed calipers, turn Cursor Select.
NOTE: After you complete the measurement, to erase all data that has been measured to this point,
but not data entered onto worksheets, press Clear.
4-4-13-2 Check Circumference/Area (Ellipse) Measurement
1.) Press Measure once to display an active caliper.
2.) Move the Trackball to position the active caliper.
3.) Press Set to fix the start point.
4.) The system fixes the first caliper and displays a second active caliper.
5.) Move the Trackball to position the second caliper.
6.) Turn the Ellipse control; an ellipse with an initial circle shape appears.
NOTE: Be careful not to press the Ellipse control as this activates the Body Pattern.
7.) Move the Trackball to position the ellipse and to size the measured axes (the calipers).
8.) To increase the size, turn the Ellipse control in a clockwise direction. To decrease the size, turn the
Ellipse control in a counterclockwise direction.
9.) To toggle between active calipers, press Measure.
10.)Press Set to complete the measurement.
11.)The system displays the circumference and area in the measurement results window.
NOTE: Before you complete a measurement:
To erase the ellipse and the current data measured, press Clear once. The original caliper is
displayed to restart the measurement.
To exit the measurement function without completing the measurement, press Clear again.
4-4-13-3 Worksheets
Measurement/Calculation worksheets are available to display and edit measurements and calculations.
There are generic worksheets as well as Application specific worksheets. The worksheets are selected
from the Measurement Touch Panel.

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals