GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-12 - Configuring Connectivity 3 - 79
3-12-8 Instructions for setting up Network Storage on LOGIQ S8 R4
1.) IT department will need to create a share folder on the destination PC and provide the name of the
Network Share Folder, Static IP Address, Password, User Name and Name of the Shared
Directory.
Enter a Name for the Network Share and the IP address provided by IT.
Select Utility >Connectivity on the touch screen then select the Device Tab.
On the Device Tab, click “Add” enter New Device Name and IP Address. Click Save. Select Ping
to verify the icon changes (smiley face appears).
2.) Select the Service Tab. The Destination Device displayed should be the Device that was created
in the previous step, if not use the drop down arrow to select it.
Select Service Type to Add, click the drop down and select Network Storage, then click “Add.”
For Properties, enter the Password, User Name and Shared Directory provided by the IT
department for the Share that was created on the customer's computer, and then click Save.

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals