IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
4 - 40 Section 4-4 - Functional Checks
4-4-17-6 Gel Warmer Test
1.) Insert Gel Bottle into Gel Warmer.
2.) Power ON the warmer.
Figure 4-29 Gel warmer ON/OFF switch
3.) Verify Gel gets warm.
NOTE: Depending on environment, Gel requires 30 minute or more to warm up.
4-4-17-7 Single CWD
1.) Connect P2D or P6D probe. Select cardiac application like “Adult”. And start scan.
2.) Verify that CW timeline image appears.