GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
6 - 4 Section 6-2 - Monitor Adjustment
6-2-1-1-3 Gamma
1.) Press the MENU (middle) key of the monitor controls over 10 sec.
2.) Select the Picture -> Color Temp -> GAMMA by pressing the < LEFT or RIGHT > button to move
the cursor and the MENU (middle) key.
3.) Select 2.2 or 2.4 by pressing the MENU (middle) key.
6-2-1-1-4 Mode
1.) Press the MENU (middle) key of the monitor controls over 10 sec.
2.) Select the Picture -> Color Temp -> MODE by pressing the < LEFT or RIGHT > button to move the
cursor and the MENU (middle) key.
3.) Select desired value by pressing the MENU (middle) key.
4.) If selecting USER mode, adjust the R/G/B value by pressing the < LEFT or RIGHT > button to
decrease/increase the value.