EasyManua.ls Logo

GE LOGIQ S8

GE LOGIQ S8
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
7 - 38 Section 7-5 - Gathering Troubleshooting Data (R4.2.5x and later)
Section 7-5
Gathering Troubleshooting Data (R4.2.5x and later)
7-5-1 Purpose of this section
Problem images and system data (logs) can be acquired at the LOGIQ™ S8 or through service remote
connectivity. Use this data to perform service at the LOGIQ™ S8 or to send it back to the manufacturer
for analysis.
NOTE: Login before gathering log files.
To gather log files, the system requires a user to be logged in. The error message below appears when
hitting Alt+D key without any user logged in.
7-5-2 Collecting vital system information
The following information is necessary to properly analyze data or images being reported as a
malfunction or a LOGIQ™ S8 being returned to the manufacturer:
NOTE:This information is normally collected with Alt+D or Gather Logs.
Product Name = LOGIQ™ S8
Navigate to Utility > System > About.
Under Software, record:
- Software Version
- Software Revision
- Software Part Number
- Build Name
- Build View
- Build Date
Under System Image, record:
- Image Part Number
- Image Date

Table of Contents

Other manuals for GE LOGIQ S8

Related product manuals