GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
7 - 38 Section 7-5 - Gathering Troubleshooting Data (R4.2.5x and later)
Section 7-5
Gathering Troubleshooting Data (R4.2.5x and later)
7-5-1 Purpose of this section
Problem images and system data (logs) can be acquired at the LOGIQ™ S8 or through service remote
connectivity. Use this data to perform service at the LOGIQ™ S8 or to send it back to the manufacturer
for analysis.
NOTE: Login before gathering log files.
To gather log files, the system requires a user to be logged in. The error message below appears when
hitting Alt+D key without any user logged in.
7-5-2 Collecting vital system information
The following information is necessary to properly analyze data or images being reported as a
malfunction or a LOGIQ™ S8 being returned to the manufacturer:
NOTE:This information is normally collected with Alt+D or Gather Logs.
Product Name = LOGIQ™ S8
Navigate to Utility > System > About.
Under Software, record:
- Software Version
- Software Revision
- Software Part Number
- Build Name
- Build View
- Build Date
Under System Image, record:
- Image Part Number
- Image Date

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals