EasyManuals Logo

GE LOGIQ S8 User Manual

GE LOGIQ S8
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #350 background imageLoading...
Page #350 background image
GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
7 - 80 Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
7-7-16-1 Note for FIbroscan
1.) Sleep mode
FibroScan may not start after sleep mode because the time for booting the FibroScan is not enough.
Do not use sleep mode when using FibroScan.
2.) Pointint cursor
Pointing cursor is available for only the following red area of the screen during FibroScan Mode.
3.) Measurements in Report/Worksheet
Order of the measurements isn't equal between FibroScan Report and LS8 Report/Worksheet.
However, measurements results are correct.
7-7-16-2 Issues
1.) CAP and KPa values occasionally not displayed.
At the 1st time of FibroScan mode after system starts, CAP score and Fibrosis KPa values
occasionally not displayed.
<RECOVERY>
Once exit from FibroScan mode, and enter to FibroScan mode again.

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals