DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
9 - 32 Section 9-8 - Electrical Safety Tests
9-8-11 Mains on applied part (cont’d)
Where applicable, a typical test setup of non-TEE Probes can be as illustrated in 9-8-10 "Probe leakage
current test" on page 9-28.
A typical test setup for TEE probes could be as indicated below:
Figure 9-14 TEE Probe Leakage Isolation (Sink) Current Test
The test passes when the reading measure less than the values in: 9-13 on page 9-19.
WARNINGWARNING
The handle of the TEE probes must not be immersed