Revision L MAC™ 3500 Resting ECG Analysis System 5-39
2021337-036
Maintenance: Functional Checkout Procedures
Ground Continuity Resistance
1.
AC mains power cord ground prong to exposed metal
surface (ground lug)
N/A
______
Pass/Fail
Less than
200
m
1. NC = Normal Condition
SFC = Single Fault Condition
N/A = Not Applicable
2. UUT = Unit Under Test
3. All SIPs/SOPs grounded.
Table 6. Electrical Safety Checks
Step
1
Condition
2
UUT - ON
Result
Leakage
Current Limits