4-28 MAC™ 3500 Resting ECG Analysis System Revision L
2021337-036
Troubleshooting: System Errors
System Errors
The following errors may occur while you are operating this system. You may be
required to perform some action.
If you perform the recommended actions an
d the condition still remains, contact
authorized service personnel. See “How to Reach Us” to find out how to contact GE.
NOTE
For information about troubleshooting the MobileLink Standard Security
opt
ion, see “MobileLink Installation & Troubleshooting Guide” (PN 2002783-
060).
For information about troubleshooting the MobileLink Ultra High Security
opt
ion, see “MobileLink UHS Installation & Troubleshooting Guide” (PN
2020300-051).
Problem Cause Solution
appears on the screen.
No battery is installed in the system.
Install a battery and connect the system to
an AC wall
outlet to charge the battery.
flashes intermittently.
The battery charge is low.
Connect the system to an AC wall outlet to
charge the battery.
appears on the screen.
The writer door is open.
Close the writer door.
The system does not power up when
operating from battery power.
The battery is empty.
Connect the system to an AC wall outlet to
charge the battery.
The system shuts down when operating
from batte
ry power.
Battery is empty, or the Automatic Shutdown
feature is enabled.
Connect the system to an AC wall outlet to
charge the battery, or power on the system.
“X” Lead disconnected message appears. Electrode(s) disconnected. Reconnect the electrode(s).
MODEM ERROR. The remote device is
not responding. Would you
like to retry?
Modem not connected. (If wireless option,
client bridge not connected.)
Connect and retry.
(Wireless option only) MAC 3500 is not
wit
hin range of access point.
Relocate MAC 3500 to within range of
access point and retry transmission.
Cannot use the system because D
evice
Password does not work.
Device Password has been changed or has
not been adequately communicated to the
staff.
Override the Device Password prompt by
pressing the following keys at the same time:
63A, 64A, 65A
i

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 3500 and is the answer not in the manual?

GE MAC 3500 Specifications

General IconGeneral
BrandGE
ModelMAC 3500
CategoryMedical Equipment
LanguageEnglish

Related product manuals