B-4 MAC™ 3500 Resting ECG Analysis System Revision L
2021337-036
Software/Hardware Compatibility: Circuit Board Compatibility Matrix
Circuit Board Compatibility Matrix
Not all supported software versions are compatible with all CPU/Comm board
combinations. You must know which software versions are compatible with which
CPU and Comm boards before you replace either board or upgrade your software.
The following table identifies which software
versions are compatible with which
CPU/Comm board combinations.
Table 3. Supported Hardware/Software Combinations
Main Board
Primary Board
Bo
ot Code
Software
Application
Comm Boards
-002 -003 -004
-008 B4
10B
9D
-007 B3
10B
10A
1
9D
9C
2
-006 B3
10B
10A
1
9D
9C
2
1
The 2022332–004 Comm Board is not compatible with the version 10A
software. If you need to upgrade to this comm board on a device with version
10A software, you must upgrade to software version 10B.
2
The 2022332–004 Comm Board is not compatible with the version 9C
software. If you need to upgrade to this comm board on a device with 9C
software, you must upgrade to software version 9D.