Revision L MAC™ 3500 Resting ECG Analysis System 5-13
2021337-036
Maintenance: Keypad Replacement
NOTE
The eight plastic standoffs should remai
n with the keyboard. However, if
any do remain in the top cover, use a small pliers to extract from the top
cover. Do not damage the rubber cover which is permanently affixed to the
top cover.
5. Using a small pliers, remove the eight plastic standoffs that remained in the top
cover.
6. Place new keyboard in place on top cover. B
e sure you align the eight plastic
standoffs with
the appropriate holes in the top cover.
7. Once the keyboard has been properly aligned with the top cover, push down on
keyboard at each of the eight s
tandoff locations until it snaps into place on
the
to
p cover.
Reverse the procedures for removal of the keyb
oard, replacing all fasteners
whi
ch were removed previously.