2.5 Magnetic Resonance Elastography (MRE) Option
2.5.1 Requirements
1. Customer must work with the RF shield vendor to provide a waveguide for the 1 in. (25.4
mm)
2.
MRE Resoundant Acoustic Driver location is limited to the length of the 1 in. (25.4 mm) tube
(see the usable cable lengths in Chapter 6, Section 1.4, Magnetic Resonance Elastography
(MRE) Option)
3. The MRE Resoundant Acoustic Driver module requires a 110/220 VAC, 50/60 Hz, 2.0 A
facility supplied outlet. Power at the outlet must be continuously available
2.5.2 Specifications
1. Weight: 53.4 lbs (24.22 kg)
2.
Gauss Limit: 50 gauss
3. Power Cord Length:
60 Hz: 240 in. (6096 mm)
50 Hz: 300 in. (7620 mm)
Illustration 4-10: Magnetic Resonance Elastography (MRE) Resoundant Acoustic Driver