1.3 Long / Short cable selection guidance
This section describes the Long / Short cable selection guidance referring to Illustration 6-1.
NOTE:
This guidance is based on the assumption that the distance between cable tray and
floor is 8.53 feet (2.6m).
Magnet Room Cable
If the Magnet – ISC Distance is less than 17.7 feet (5.4 m) , select short cable.
If the Magnet – ISC Distance is in between 17.7 feet (5.4 m) and 30.8 feet (9.4 m) , select
long cable.
Equipment Room Cable
If the ICC – ISC Distance is less than 1.3 feet (0.4 m) , select short cable.
If the ICC – ISC Distance is in between 1.3 feet (0.4 m) and 21 feet (6.4m) , select long
cable.
Illustration 6-1: Distance for cable selection
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
Chapter 6 System Interconnections 131