Illustration 6-6: Cable Tray Requirements (90° Magnet Interface)
2.3 Equipment Room Requirements
1. All equipment interconnects must route overhead
2.
Cables/hoses must drop through the bottom or off the end of the cable support directly to
the top of the cabinets
3.
Cable tray must be above minimum ceiling height (15.7 in. (400 mm) above the ISC
cabinet).
4.
Excess cable length must be stored in the Equipment room
5.
For multiple MR system installations, cables from different MR systems must not share the
same cable support
6. The table below lists the minimum width for cable trays between Equipment Room (and
Operator Workspace) components
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
138 2 MR System Interconnects Routing Requirements