1.2 Usable Cable Lengths
Four configurations of cable lengths are available for order. Measure the distances between the
equipment and compare them to the distances specified in the tables. If the your distance is less
than the short distance specified, the Site Option is “Short”; otherwise, the Site Option is “Long.”
Table 6-3: Order Configuration Options
Configuration Equipment Room — Site Option Magnet Room — Site Option
A Short Short
B Long Short
C Short Long
D Long Long
Table 6-4: Usable Cable Lengths
Point A Point B
Site Option: Short
Site Option: Long
in. (mm)
Equipment Room
CRY Magnet Monitor 370.1 (9400)
Magnet Monitor ICC 661.4 (16800)
ISC, top panel Magnet Monitor 645.7 (16400)
ISC, top panel GOC, rear panel 97.5 (25000)
ICC, top panel GOC, rear panel 97.5 (25000)
ISC, top panel ICC, top panel 15.7 (400) 252 (6400)
ISC, top panel Magnet room door switch 960.6 (24400)
ICC, Pen Panel
E-off switch, Control room or
Equipment room 661.4 (16800)
Magnet Room
ISC Pen Panel
MAG, gradient cable clamp
block 210.6 (5400) 366.6 (9400)
ICC Pen Panel MAG 511.8 (13000)
ICC Pen Panel E-off switch, Magnet Room 1173.2 (29800)
MRU
MAG, gradient cable clamp
block 972.4 (24700)
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
130 1 MR System Interconnects Specifications