Component Parameter Requirements
Power Availability Continuous, facility power is required at all times for operation of the Cryocooler
(CRY) to minimize cryogen consumption.
Service re‐
ceptacle in
Magnet
Room
Voltage / Frequency
100-120 VAC 50/60 Hz Receptacle required for small power tools. Local volt‐
age and portable transformers for voltage values.
Phase 1
Maximum Amps 2.0
Table 2-13: System Power Demand
Equipment Power Draw (kVA)
ISC PDU Continuous Power 55
ISC PDU 5 Second Power 65
Cryo Compressor Continuous Power 9
Total System 5 Second Power 77
Total system Continuous Power 64
Standby Consumption (no scan) (PDU, CRY) < 17
Table 2-14: Power Consumption
Power Consumption Sleep Mode 5.7kW
Power Consumption Standby(no scan) 11.1kW
Power Consumption Typical consumption 16.1kW
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
50 10 MR Suite Electrical Requirements