b. Terminal blocks to accept 12 AWG (E0009) on the Cryocooler
c. Terminal blocks to accept 1 or 2 AWG on the PDU in ISC
8.
The optional GE Healthcare MDP consists of the following:
a.
A 3-pole 200A main circuit breaker rated for the total current of all the sub-breaker
circuits
b.
A 3-pole 150A circuit breaker rated for the current of the PDU circuit
c.
A Customer Supplied 3-pole 30A circuit breaker is required for Cryocooler circuit outside
of MDP
d. All circuit breakers have a short circuit current interrupting rating of 25000 Amps
e.
Auto restart on the Cryocooler circuit
f. Emergency off circuit including 2 pushbuttons to be installed external to the MDP
g. Terminal blocks that can accept 3/0 AWG maximum size on the main, 8 AWG maximum
size on the 30A Customer Supplied Breaker for Cryocooler circuit, and 1 or 2 AWG on
the PDU in ISC.
NOTE:
12 AWG (Cable RUN # E0009) will be connected between the 30A Customer
Supplied Breaker and the Cryocooler.
h. Isolated neutral terminal block for termination
i. Multiple ground terminal blocks as required by panel design
j.
Listed and labeled by a Nationally Recognized Testing Lab (NRTL) in accordance with
UL 508A and IEC/EN 60204–1 and bear the CE Marking in accordance with the EU Low
Voltage Directive (2006/95/EC) and Electromagnetic Compatibility Directive (2004/108/
ED).
k.
Power on indicators
l. Two isolated, normally open contact pairs that open when e-OFF is pressed for use with
optional accessories
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
Chapter 2 General System Level 53