10. The system must have a manual exhaust fan switch near the Operator Workspace (OW)
and in the Magnet room near the door (the switches must be connected in parallel)
NOTE:
If the Magnet room contains an optional oxygen monitor, the Magnet room switch
is not required.
11.
All system components must be accessible for customer inspection, cleaning, and
maintenance
12.
Electromechanical fire dampers must not be used. Fusible link fire dampers may be used
(with annual inspection)
Illustration 3-13: Magnet Room Exhaust Fan Schematic
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
84 8 Magnet Room Venting Requirements