ActionStep
Adjust the probe connectors of Plus or Plus Flex bag assembly by gently
pulling the connectors through the probe window.
8
Adjust impeller position
If the impeller baseplate is not correctly installed, the agitator system is not properly
magnetically coupled and does not function correctly. If the agitator emits a loud clicking
noise during operation, the impeller baseplate is decoupled from the magnetic drive
head.
To correct the impeller position, follow the instructions below.
ActionStep
Lift the bag upwards and disengage the impeller baseplate from the agitator
motor.
1
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 153
6 Preparation
6.3 Prepare the disposable bag
6.3.2 Install the disposable bag