ActionStep
Install two ratchet clamps over the section of tubing between the bag as-
sembly probe port (welded into the disposable bag) and the female aseptic
connection device (ACD) connector attached to the probe port of the dispos-
able bag, at the location shown in the following illustration.
1
Using your fingers, loosely tighten the ratchet clamps until the teeth on the
clamp are engaged enough to hold the clamp in position (2 to 3 teeth en-
gaged).
2
NOTICE
Do not tighten the clamps too much, otherwise they
prevent the insertion of the probe into the disposable
bag.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 169
6 Preparation
6.4 Prepare and install the probes
6.4.4 Insert probes into the disposable bag