ActionStep
Attach the temperature probe cable to the probe support bar (for example
with a zip tie) to prevent operators from accidentally removing the temper-
ature probe from the thermowell.
4
Put the temperature probe cable into the cable holders on the short side of
the mixer vessel.
5
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 181
6 Preparation
6.4 Prepare and install the probes
6.4.4 Insert probes into the disposable bag