ActionStep
On the Prime Pumps screen, change the pump direction for Pump 1 (Acid
pump, 1), or for Pump 2 (Base pump, 2) by dragging the toggle button
to CW or CCW.
3
1
2
See Tubing in the pump, on page 184 for more information about the liquid
flow direction through the pump.
Tap BACK to return to the pH Detail screen.4
Prime the pump
Follow the instructions below to prime the pump.
ActionStep
Make sure the pH probes are installed in the disposable bag. See Section 6.4
Prepare and install the probes, on page 157 for more information.
1
212 Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
7 Operation
7.3 Operate Xcellerex XDUO mixer
7.3.3 Prepare the pump