Line Sub-
line
Parameter Basic
setting
Setting range
Authorization
Row 96: Activation inputs
96 01 Activation, level A 1
0=input not active, 1=input active
CD
96 02 Activation, level B 1
0=input not active, 1=input active
CD
96 03 Activation, level C 1
0=input not active, 1=input active
CD
96 04 Activation, descaler, low level 1
0=input not active, 1=input active
CD
96 05 Activation descaler empty 0
0=input not active, 1=input active
CD
96 06 Activation chemical agent low level 0
0=input not active, 1=input active
CD
96 07 Activation chemical agent empty 0
0=input not active, 1=input active
CD
96 08 Activation rinse-aid low level 0
0=input not active, 1=input active
CD
96 09 Activation rinse-aid empty 0
0=input not active, 1=input active
CD
96 10 Activation, process agent, low level 0
0=input not active, 1=input active
CD
96 11 Activation, process agent, empty 0
0=input not active, 1=input active
CD
96 12 Activation spray agent low level 0
0=input not active, 1=input active
CD
96 13 Activation spray agent empty 0
0=input not active, 1=input active
CD
96 14 Activation, leakage sensor 1
0=input not active, 1=input active
CD
96 15 Activation, door closed 1
0=input not active, 1=input active
CD
96 16 Activation, door closed 1
0=input not active, 1=input active
CD
96 17 Door open (FD1810) 1
0=input not active, 1=input active
CDE
96 20 Safety switch door closing (FD1810) 1
0=input not active, 1=input active
CDE
96 21 Pedal (FD1810) 0
0=input not active, 1=input active
CDE
Line 97: Activation outputs
97 01 Activation, pressure pump 1
0=input not active, 1=input active
CD
97 02 Activation, steam generator 1
0=input not active, 1=input active CD
97 03 Activation, cold water 1
0=input not active, 1=input active CD
97 04 Activation, hot water 1
0=input not active, 1=input active
CD
97 05 Activation, soft water 0 0=input not active, 1=input active CD
97 06 PE system (rotary nozzle) 0
0=input not active, 1=input active CDE
97 07 SP system (spray nozzle) 0
0=input not active, 1=input active CDE
97 08 Extra nozzle 0
0=input not active, 1=input active CDE
97 10 Activation, pump, descaling 1
0=input not active, 1=input active CD
97 11 Activation pump chemical agent 0
0=input not active, 1=input active CD
97 12 Activation pump rinse-aid 0
0=input not active, 1=input active CD
97 13 Activation, pump, process agent 0
0=input not active, 1=input active CD
97 14 Activation pump spray agent 0
0=input not active, 1=input active
CD
97 15 Activation, extra element, steam
generator
1
0=input not active, 1=input active
CD
97 16 Activation, condensate cooler 0
0=input not active, 1=input active
CD
97 17 Activation, external cooling. 0
0=input not active, 1=input active
CD
97 23 Activation relay corresp. to yellow LED 0
0=input not active, 1=input active
CD
97 24 Activation relay corresp. to red LED 0
0=input not active, 1=input active
CD
97 25 Rim flushing 0
0=input not active, 1=input active
CDE