micro balance Mod. E50S/3
Page 14 of 38
6.3.5 Calibration functions
SET
MODE OFF
Closed
EXTERNAL
External
EXT.-DEF.
External with user-defined weight (“DEF. n.nnn g")
INTERNAL
with internal weight
AUTO
Automatic (AUTOCAL)
DEF. 0.000 g
Calibration weight for EXT.-DEF. mode
AUTOCAL. TIME/TEMP.
Auto-calibration on time and temperature
TEMPERATURE
Auto-calibration on temperature
TIME
Auto-calibration on time
AUTOCAL-TIME
6 h
Time for auto-calibration
For calibration of the balance see chapter 5.7 "Calibration of the balance" and see chapter 12.3 "Calibration".
6.3.6 Weighing mode
SET WEIGHING
FLOATINGDISPLAY 0.08
Input integration time (in seconds)
0.16
0.32
STABILITY LOW
Setting the stability control (instability of the
balance location)
MEDIUM
HIGH
AUTO-STANDBY OFF
Auto-Standby not active or active after nn minutes
0.5 MIN.
1 MIN.
5 MIN.
10 MIN.
AUTO-ZERO ON/OFF
Automatic zero correction on/off
QUICK-TARE ON/OFF
Quick-Tare on/off
With the aid of the Weighing mode functions, you may define the quality of the balance location. (see chapter
5.4 "Choosing a suitable location")
With the aid of the “AUTO-STANDBY” function, you can define the period of non-use before the balance
automatically switches into the energy-saving mode.
NOTE
The Auto-Standby function only works with the automatic zero-correction activated (“AUTO-ZERO”).
“FLOATINGDISPLAY”
The value set for “FLOATINGDISPLAY” defines the period after which each new measurement is displayed.
For the definition of this period, the quality of the balance location is crucial. The stability control must also
be set appropriately.
Recommended values:
Optimum balance location: “FLOATINGDISPLAY 0.08”
Good balance location: “FLOATINGDISPLAY 0.16”
Critical balance location: “FLOATINGDISPLAY 0.32”