Semi-micro balance Mod. E50S/3
Page 15 of 38
NOTE
The value of the Floating Display is a function of the stability control and the balance location. For
balance location, see chapter 5.4 "Choosing a suitable location" and see chapter 6.3.6 "Weighing
mode".
“STABILITY”
The value set for the stability control depends on the quality of the balance location and must be correctly
chosen in order to obtain optimum, reproducible results. Choose:
Optimum balance location: “STABILITY HIGH”
Good balance location: “STABILITY MEDIUM”
Critical balance location: “STABILITY LOW”
“AUTO-STANDBY”
The Auto-Standby mode turns off the balance automatically, if:
the balance is tared and has shown .Zero. for at least 5 minutes
the balance has received no remote control command via the interface for at least 5 minutes,
the automatic zero correction .AUTO-ZERO. is activated.
It is possible to re-start the balance after it has been switched off by an automatic Auto-Standby:
Briefly press any key
Put a weight on the balance
Make a remote control command via the interface
“AUTO-ZERO”
If the automatic zero correction .AUTO-ZERO. is activated, the balance always gives a stable zero (e.g. even
with fluctuations in room temperature).
6.3.7 Interface functions
SET
BAUDRATE 300
Select baud rate
600
1200
2400
4800
9600
19200
PARITY 7-EVEN-1STOP
Select parity
7-ODD-1STOP
7-NO-2STOP
8-NO-1STOP
HANDSHAKE
NO
Select handshake function
XON-XOFF
HARDWARE
The RS232/V24 interface on the device is matched to the interface of a peripheral device with the aid of the
interface functions. (see chapter 11 "Data transfer")
6.3.8 Date and time
SET
DATE [DD.MM.YY]
Set date and time
TIME [HH.MM.SS]
FORMAT STANDARD/US