CHEMICAL LEVEL COMPONENT CYCLES* CONDITIONS
STERIS
®
eSSENTIALS
Concentrates
Cleaner Blade 750
Water temperature: 3040°C (86–104°F)
Exposure: Prepare solution at 0.1251ounce/gallon (1–8mL/L).
Soak component for 5 minutes. Before removing from solution,
brush all surfaces and pay special attention to hard-to-reach
areas. Use a syringe to flush the connector.
Rinse: Rinse for 3 minutes under running water. Flush the
connector with running water and a syringe.
Pro-Line
Solutions
EcoZyme
®
Cleaner
Blade 3000
Water temperature: 1929°C (6684°F)
Exposure: Prepare solution at 1ounce/gallon (7.81 mL/L) in
3040°C (8610F) water. Soak component for 5 minutes.
Before removing from solution, brush all surfaces and pay special
attention to hard-to-reach areas. Using a syringe, flush the
connectors.
Rinse: Rinse for 5 minutes under running water. Using a syringe,
flush the connectors.
Video Cable 3000
Smart Cable 1500
Metrex
®
CaviWipes™
Cleaner
Video Cable 1500
Water temperature: N/A
Exposure: Use towelette(s) to remove all visible contamination
from the component. Using fresh towelette(s), wet all surfaces
and allow to remain wet for 3 minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Smart Cable 1500
Wip’Anios
Premium
Cleaner
Video Cable 1500
Water temperature: N/A
Exposure: Use towelette(s) to remove all visible contamination
from the component. Using fresh towelette(s), wet all surfaces
and allow to remain wet for 5 minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Smart Cable 1500
Clinell
®
Universal
Sanitizing
Wipes
Cleaner
Video Cable 1500
Water temperature: N/A
Exposure: Use towelette(s) to remove all visible contamination
from the component. Using fresh towelette(s), wet all surfaces
and allow to remain wet for 5 minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Smart Cable 1500
Sani Cloth
®
Active Wipes
Cleaner
Video Cable 1500
Water temperature: N/A
Exposure: Use towelette(s) to remove all visible contamination
from the component. Using fresh towelette(s), wet all surfaces
and allow to remain wet for 5 minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Smart Cable 1500
* Value indicates number of compatibility cycles tested on the component. Exceeding the recommended number of cycles may affect the potential life of
the product.
Do not use this solution to clean a GlideScope Smart Cable.