CS510/IS210-UM-251-9370 8-15
8. DISASSEMBLING AND ADJUSTING THE MECHANICAL PARTS
8.15 CIS (Charge Coupled Device Imaging Sensor) Board
Detaching the CIS boards
(1) Detach the contact glass assembly (see Section 8.13).
(2) Detach each CIS assembly (see Section 8.14).
(3) Remove the two M2L5 binding head screws supporting each CIS board.
M2L5 binding head screwM2L5 binding head screw
CIS boardCIS board
(4) Disconnect the cables from each CIS board and then detach each CIS board.
Reattaching the CIS boards
(1) Reattach each CIS board in the reverse order in which it was detached.
(2) Prepare corresponded CIS board before replacing the CIS board.
(3) When you have replaced the CIS boards, perform calibration using the software adjustment.
When replacing the
CIS boards
, be sure to clean the interior of the image sensor unit and the inside of the contact
glass assembly. A soiled or dusty unit can affect scanning performance.