CS510/IS210-UM-251-9370 10-7
10. ADJUSTMENTS USING THE SOFTWARE
10.4 Making Adjustments
When you replaced the main board or the CIS sensors, follow the sequence of adjustments below.
1. Calibration (white correction)
2. Feed distance adjustment
3. Sensor stitch alignment
5. Recalibration (white correction)
4. Scanning start position adjustment
6. Color correction
Before beginning calibration, clean the contact glass plates and scanning-table surface. Any dust or dirt
on those surfaces may affect the calibration results and resulting image quality.
Check that the calibration sheet is free of any dust or dirt.