Menu Structure & Configuration
36
To change the name of the unit, press on the grey box beside Unit Name, a pop-up keyboard
will appear for you to enter the designed texts and numbers.
Beam path
The beam path is the distance between the transceiver and the reflector in cm. Value is
currently not in use.
Alarm level
The alarm level can be freely configured upon the requirement of each applica- tion. Note
that the monitor and control unit will always provide a warning which is 50% of the Alarm
level. This warning can be used as lens-cleaning warning
Alarm delay
Alarm delay is the delay time in second for the alarm to be activated.
Alarm group
The system has two alarm groups: Group 1 and Group 2. The alarm outputs are configured
as follows:
Thus, by setting Alarm Group for each sensor, you configure which alarm outputs are con-
nected to that sensor.