38 Model 675 User Instructions
6. Key in the load cell's mV/V value and press [ENTER].
- or -
6a. Press [ENTER] to accept the displayed value.
7. Steps 5-6 will be repeated for as many load cells as specified in step 4.
8. The display prompts LC FS showing the value last entered for the load cell full
scale.
9. Key in the full scale capacity for the load cell(s) and press [ENTER].
- or -
9a. Press [ENTER] to accept the displayed value.
10. The display briefly shows Updtg Gains as it updates the gain values, then
prompts CurWt Zero?.
11. Press [ENTER] to establish the current input signal as the zero reference.
- or -
11a. Press [SELECT] to display Zero=0mVv?. Press [ENTER] to use a 0mV/V
output as the zero reference.
- or -
11b. Press [SELECT] to display Keyin CurWt. Key in the known gross weight
already applied to the scale and press [ENTER].
- or -
11c. Press [CLR] to bypass the zeroing option.
12. The display shows CAL OK? suggesting that the calibration is acceptable.
Accept the calibration by pressing [ENTER] at the CAL OK? prompt.
- or -
12a. Repeat the calibration by pressing [CLR] at the CAL OK? prompt.
13. Once the calibration is accepted in step 5, press [ENTER] at the
ENTER=SAVE prompt and again at the ENTER=EXIT prompt to save the new
calibration and exit the calibration mode.
- or -
13a. To exit the calibration mode without saving the new calibration, press [CLR] at
the ENTER=SAVE prompt. Then press [ENTER] at the ENTER=UNDO
prompt and again at the ENTER=EXIT prompt to exit the calibration mode.
4.3.7 Multi - Scale Calibration
When more than one scale is enabled, the prompt Keyin Scl# appears before
accessing the calibration method selections. Key in the scale number to be calibrated
and press [ENTER]. Proceed with a calibration method as described in Calibration
Methods on page 33. After completing a calibration, the Keyin Scl# appears once
again. Enter the next scale number to be calibrated, or press [CLR] to exit the
calibration mode and save the new calibration data.