Model 675 User Instructions 41
5.3 ICalibration Errors
5.4 General Errors
Message Description
CoDE30
F.S.>
MAX!
The entered calibration weight, together with the currently applied signal, indicates
that the full scale signal will be greater than the allowed maximum of the controller.
CoDE31
F.S.<
.1mVv
The entered calibration weight, together with the currently applied signal, indicates
that the full scale signal will be less than the allowed minimum of the controller.
CoDE32
ADD
MORE!
The applied weight during calibration was less than 0.1% of capacity. More weight
than this is required.
CoDE33
ReCAL
Req'd
The just completed calibration is insufficient to guarantee accurate results due to
either the cal weight being less than 5% of capacity or this was the first calibration of
this platform to this Indicator and, therefore, the coarse gain was adjusted by the
Indicator.
CoDE39
"A/D˜
Cal
The A/D calibration values for one of the scales have not been entered.
Message Description
CoDE41
Idnot
Used!
If a string type variable that has not been setup (for example, had its size set to a non-
zero value.) or if no string type variables have been setup.
CoDE42
Check
Setup
A sample operation was performed with a sample size of "0" at parameter P34.
CoDE43
dbNOT
SETUP
An attempt to access an undefined database occurred.
CoDE51
Too
Small
The sample placed on the platform is too small to accurately compute the piece
weight. Increase the sample size.
CoDE52
Can't
Count
There is an insufficient quantity on the platform to perform an accurate count.
CoDE53
Accy<
Req'D
The accuracy is less than required. Increase the sample size.
CoDE54
Scale
Disbl
An attempt to select a disabled scale was made.
CoDE57
dbOpt
Error
The FRAM database is corrupt (valid signature, invalid checksum). Contact GSE.
CoDE60
New
FLASH
A new FLASH file has been uploaded via the ReFlash or BDM flash utilities in order to
upgrade the firmware. Press any key to acknowledge.