EasyManua.ls Logo

GW Instek ASR-2000 Series - [:Source]:Simulation:initial:voltage; [:Source]:Simulation:normal<1|2>:Code

GW Instek ASR-2000 Series
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
101
[:SOURce]:SIMulation:INITial:VOLTage
Set
Query
Sets or queries the Vset parameter of the initial
step for the Simulation mode.
[:SOURce]:SIMulation:INITial:VOLTage
{<NR2>|MINimum|MAXimum}
[:SOURce]:SIMulation:INITial:VOLTage?
[MINimum|MAXimum]
<NR2>
Voltage of the initial step.
MINimum
Minimum settable voltage
MAXimum
Maximum settable voltage
:SIM:INIT:VOLT MAX
Sets the initial step voltage to the maximum.
[:SOURce]:SIMulation:NORMal<1|2>
:CODE
Set
Query
Sets the external trigger output for the normal 1 or
normal 2 step parameter. This option is only
applicable when in the Simulation mode.
[:SOURce]:SIMulation:NORMal<1|2>:CODE
{<NR1>|MINimum|MAXimum}
[:SOURce]:SIMulation:NORMal<1|2>:CODE?
[MINimum|MAXimum]
<NR1>
0=LL, 1=LH, 2=HL, 3=HH
MINimum
0 (LL)
MAXimum
3 (HH)
SIM:NORM1:CODE 1
[:SOURce]:SIMulation:NORMal 1
:FREQuency
Set
Query
Sets or queries the frequency of the normal1 step of
the simulation mode.

Table of Contents