The accuracy does not match the specification.
How frequently should the power source be calibrated?
Is it proper to combine 2 or 3 units to reach 1P3W or 3P4W output?
The accuracy does not match the specification.
Make sure the device is powered On for at least 30 minutes, within
+18°C~+28°C. This is necessary to stabilize the unit to match the
specification.
How frequently should the power source be calibrated?
The ASR-3000 should be calibrated by an authorized service center
at least every 2 years.
For details regarding calibration, contact your local dealer or
GWInstek.
Is it proper to combine 2 or 3 units to reach the 1P3W or
3P4W output?
ASR-3000 doesn’t support 3P4W output function on the basis of 3
units combination. However, it supports 1P3W output type via 2
units combination in series with EXT Mode setup.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com / marketing@goodwill.com.tw.