GETTING STARTED 
31 
Over Power protection for power stage board. OPP 
is a software protection function that is 
corresponded with VA value. When the unit has 
loaded less then protection point that the alarm be 
cleared. 
Sense alarm. This alarm will detect if the sense 
wires have been connected to the wrong 
polarity. 
AC input failure. This alarm function is 
activated when a low AC input is detected. 
Fan failure. This alarm function is activated 
when the fan RPMs drop to an abnormally low 
level. 
This alarm function is activated when 
insufficient output voltage and over 
temperature occur in PFC power stage. This 
alarm can be set by user. 
This alarm function is activated when over 
output voltage or insufficient voltage and over 
current occur in DCDC power stage. Contact 
local dealer or GW Instek directly for repair. 
This alarm function is activated when transient 
output current larger than hardware protection 
point and over temperature occur in DCAC 
power stage. This alarm can be set by user.